COLD VT
OPERATION ON THE VARIAN 500
Inverse
or BB probe on the 500 can be used from –
80o C to +
130 °C
Low
temperature VT is based on the flow rate of gas to control the “coldness”
and the heater controlling to the exact temperature.
So you “overcool” and then nudge up with the heater.
KNOW
WHERE THE EMERGENCY EJECT BUTTON IS!!
KNOW
WHERE THE EMERGENCY SWITCH ON THE HEATER IS!!
(for overheating only)
ANY
RAPID CHANGE IN TEMPERATURE CAN DAMAGE THE PROBE – DON’T!!
LOW TEMPERATURE BETWEEN - 30oC AND + 25oC using the FTS
system
If
you are going between > -30oC
and <25 °C
you may use the FTS system which is always on.
This controls with refrigeration the air coming into the probe...it cools
it down below where you want to be and then the probe heater brings it up to the
desired temperature.
1.
Put
in your sample in the normal way. Lock
and shim.
Take RT spectrum if desired. [Always keep your eye on the lock level and
the spinning in the acqstat window as this is the first sign something is going
wrong. If the sample freezes, the
lock level will drop out. If you
have moisture in the lines, the sample will not spin.]
2.
Type
“temp” on the command line, to bring up window.
Select a temperature, using the slide bar around 8
°C.
Select SP2 (
0 °C)
on the FTS unit. It should start
going to 0 °C.
3.
Set
the VT flow rate on the magnet leg box to ~ 15. DO NOT EXCEED ~ 18 or the tube
might start to lift in the probe. Watch the spinning rate as this will indicate
a problem.
4.
Type
“su” on the command line. DO NOT TYPE ANYTHING ELSE..WAIT
FOR THE BEEP .... The beep means that it is stable, and it will
say “setup complete” in the text window.
5.
You can now slowly lower the temperature and
take spectra. Lower the temperature
on the slide bar, no more than 15 degrees at a time.
You should give the sample ~10 min to equilibrate at a different
temperature, and you should re-shim. DO
NOT GO FAST!!
IF YOU ARE DOING CARBON YOU WILL NEED TO RETUNE THE PROBE FOR H
& C at YOUR FINAL TEMPERATURE - FOR PROTON RETUNING GIVES A BETTER SPECTRUM
6.
If
you need to go lower (0 to -20),
now select SP3 (
-20 °C) on
the FTS unit, and it will move to
-20. Set your new
temperature in the “temp” window; type “su” and wait for equilibration;
re-shim. To go lower in temperature (
-20 to -30) select
SP4 ( -40 °C)
on the FTS. Set the slide and wait for equilibration, then reshim.
GO SLOW! DO
THIS STEPWISE; don’t just hit SP4 and go for
it!
7.
When
you are done, set the FTS to SP1 (20
°C)
and raise the temperature
control bar by increments of 15 degrees. Take your time!
At the end, set the VT flow rate back to ~ 10
SP1 above
20 °C
SP2 0
to 20
°C
SP3
0
to -20
°C
SP4 -20
to -40 °C
LOW-TEMPERATURE – 80oC
to
+25oC
If
you are going below < - 30 °C
YOU MUST USE TANK NITROGEN and liquid nitrogen in the bucket
(be sure you have a full tank and a transport dewar with liquid
nitrogen).
FOR TROUBLE
- DROP THE LIQUID NITROGEN BUCKET FIRST !!!
Bring the temp up to 15 °C.
KNOW WHERE
THE EMERGENCY EJECT BUTTON IS !! (get
that sample out-overheating or cooling)
KNOW WHERE
THE EMERGENCY SWITCH ON THE HEATER IS !! (for
overheating)
ANY RAPID
CHANGE IN TEMPERATURE CAN DAMAGE THE PROBE – DON’T
1.
Turn
on the nitrogen tank at about 60 lbs (blue line). Turn off the air (yellow
lever). Turn on the nitrogen
(red lever) (off is perpendicular -- on is parallel to the pipe)
2.
Type
“temp” on the command line, to bring up the window.
Turn the VT OFF –
make sure it stays off.
Watch the temperature decrease until it stops around 20 °C (FTS
temp).
3.
Remove
the big black tube from the probe, and turn off the FTS (both switches).
4.
Disconnect
the tee with the incoming lines (quick disconnect) from the back of the FTS and
connect it to the input of the coil in the bucket.
5.
Check
to see if the nitrogen is flowing through the coil. Then hook up the output of
the bucket to the probe where you had removed the big black FTS connection.
6.
Hit
the “reset VT” button, then turn the temperature control “on” and slide
to select 15 °C.
7.
Check
to see that the VT flow rate on the magnet leg box is ~ 8; then wait about 5
minutes for the line and coil to flush thoroughly with nitrogen.
8.
Take
out the black rubber stopper in the white Styrofoam bucket and then pour liquid
nitrogen from a dewar into the bucket. Make sure the coil is covered.
9.
Type
“su” on the command line and wait
for the beep, when 15 °C is reached.
10.
Watch
the acqstat window. It should say
“status: VT regulation” on the top, then “setting:
15 °C”
and the “actual temperature” (temp
is also in red digital display on left of monitor).
11.
If
it goes down too far too fast, lower the VT flow rate on the gas leg a little
but no lower than five. As you
go down in temperature you should raise the flow rate back to ~10. This flow
rate controls the “coldness” you are adding.
12.
Put
in your sample in the normal way. Lock
and shim. Always keep your eye on
the lock level and the spinning in the acqstat window as this is the first sign
something is going wrong. If the
sample freezes (or boils), the lock level will drop out.
If you have moisture in the lines, the sample will not spin.
13.
You
can now slowly lower the temperature and take spectra.
Lower the temperature on the slide bar, no more than 15 degrees at a
time. Give the sample ~10 min to
equilibrate at a different temperature, and you should re-shim.
DO NOT GO FAST!!
14.
When
done, raise the temperature control bar by increments of 15 degrees, take your
time! At a steady 15 °C, lower the bucket from
the coil and turn
OFF the VT.
15.
When
the frost is gone from the coil and temperature is above 15, quickly turn off
the nitrogen tank and then the red nitrogen lever, turn the air back on with the
yellow lever.
16.
Now
reverse the procedure. Remove the
input line from the bucket to the probe.
17.
Remove
the tee from the gas input line to the bucket and put it back on the back of the
FTS system.
18.
Turn
on the FTS with the desired temperature at 20 °C (SP1).
When it is at 20 °C, reconnect the large
black line from the FTS system to the probe.
19.
Turn
on the VT to 25. Several hours later
you should wipe the Styrofoam bucket dry and put the bucket back up.
Leave the black stopper off for the rest of the day.
NEVER
TURN OFF THE CONTROLLER WITHOUT FIRST DROPPING THE BUCKET.
SHOULD YOU HAVE TO STOP IN THE MIDDLE .....DO NOT TURN OFF THE CONTROLLER.